Known problems with Smart-V-link V3.1 rev 0.0

We’re fixing the following problems confirmed with Smart-V-Link V3.1 rev 0.0 and the updater software should be available on our website toward mid-April, 2012. Please follow the instructions to avoid each problem until then:

- When PPG mode is selected on Smartdop 30EX offline with both Doppler and PPG probes connected at the same time and you go to the pressure monitoring screen, Smart-V-Link goes back to Main screen right away. => Select Doppler mode on Smartdop 30EX offline before starting pressure monitoring.- Patient information can not be entered right after installing Smart-V-Link. => Re-start Smart-V-Link and enter patient information.- When patient information is entered on Smartdop 45 and you try to download it to Smart-V-link, software freezes. => Enter patient information on Smart-V-Link.March 9, 2012